This paper deals with the determination of acrolein based on a strategy provided with a sensitive detection, the chemiluminescence; a "clean chemical" analyte treatment, the on-line UV-irradiation; and, a simple and quick automated system, the emergent multicommutation continuous-flow methodology. The pesticide solution is irradiated (stopped-flow) for 30 s; then, the resulting solution is introduced to the flow-cell after being alternated with segments of the stream formed by potassium permanganate in strong acid solution and ethanol segments. The calibration range, from 1 to 1 mg l 21 , resulted in a linear behavior over the range 5-100 mg l 21 and fitting the linear equation: I ¼ 75.69 Â 17.75; correlation coefficient 0.9997. The limit of detection was 0.1 mg l 21 and the sample throughput 52 h 21 . After testing the influence of a large series of potential interferents the method was applied to different types of samples.