Objectives : The purpose of this study is to understand the temperature characteristics depending on the thickness and material of the needle used with the Electronic Warm Acupuncture Device (EWAD). Methods : We controlled experimental environment and measured temperature changes of a silicon phantom in which K-type thermocouples were inserted at depths of 2, 7 mm. EWAD perfomed with acupuncture needles of various thicknesses (0.25×60 mm, 0.40×60 mm, and 0.50×60 mm) and materials (Gold 0.40×50 mm). We set non-needle (only heated with EWAD skin heater) group as a control group. Results : The maximum temperature and temperature changes of 0.40 mm, 0.50 mm needle group were significantly higher than the non-needle group. The highest temperature range in all needle groups was 0.50 mm needle group (41.44±0.31℃). However, the 0.25 mm needle group was not significantly different from the non-needle group. Maximum temperature of gold needle group was significantly higher than stainless steel needle group. Temperature changes of gold needle group were higher than stainless steel group at the depth of 7 mm. Conclusions : It was found that needle thickness and material of acupuncture had an effect on the temperature of the EWAD. When performing EWAD treatment, consideration of thickness and material of acupuncture is needed. Future research is needed using phantoms that can reflect actual clinical situations and better mimic the human body.