. Analytic performances were determined in three laboratories. Clinical performances were evaluated on 100 plasma samples from HIV-2-infected patients (groups A, B, and H) by comparison with the assay currently used for the ANRS HIV-2 cohort. The specificity was 100%. Sensitivity was 50 copies/ml (cp/ml) and was optimized to 10 cp/ml. The within-run coefficients of variation in the three laboratories varied from 0.54% to 1.61% at 4 log 10 copies/ml and from 7.24% to 14.32% at 2 log 10 cp/ml. The between-run coefficients of variation varied from 2.28% to 6.43%. Of the 39 clinical samples below 2 log 10 in the current assay, the new test improved the detection or quantification of 17 samples, including eight group B samples. For quantifiable samples, similar loads were obtained with the two assays for group A samples. The median difference between the two assays for group B samples was ؉0.18 but with greater heterogeneity than for group A. The HIV-2 group H sample had similar results with the two assays. This new assay is highly sensitive and accurately quantifies the most prevalent HIV-2 groups. This test will be useful for monitoring low viral loads in HIV-2-infected patients.