This paper proposed the method of removing baseline drift by eliminating local maxima such as P, R, T-wave signal region and local minima Q, S-wave signal region. We applied morphology-pair operations improved from morphology operation to the ECG signal. To eliminate overshoot in the result of morphology-pair operation, we apply median value operation to the result of morphology-pair operation. We use MIT/BIH database to estimate the proposed algorithm. Experiment result show that proposed algorithm removing baseline drift effectively without orignal ECG signal distortion.▸Keywords : ECG signal, Baseline drift, Morphology operation, Noise in ECG ⦁교신저자 : 김정홍