Aneroid Blood Pressure Manometer Calibration Rates of Devices Used in Home Health

Abstract: Hypertension is associated with cardiovascular disease, stroke, and a range of other medical sequelae. Accurate blood pressure (BP) readings, which depend on the integrity and calibration of the measuring device, are essential to identifying suboptimal BP. This study describes calibration rates of aneroid BP devices (a) utilized in home healthcare (HHC) and (b) having the needle resting within the zero accuracy indicator. BP devices from one branch of a home care agency were inspected and checked for calibrati… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2018
2018
2020
2020

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 0 publications
0
1
0
Order By: Relevance
“…Prior studies have reported 22% of aneroid gauges used in physical therapist practice are not adequately calibrated. 30 , 31 Additionally, these studies identified equipment in use that had bulb malfunction, tubing tears, and excessive wear, all of which impact the physical therapist’s ability to obtain an accurate BP reading. Regular calibration and inspection of equipment at regular intervals ranging from 6 months to 2 years is essential to assure BP measures are not under- or overestimated.…”
Section: Recommendations For Bp Screening In Outpatient Physical Thermentioning
confidence: 99%
“…Prior studies have reported 22% of aneroid gauges used in physical therapist practice are not adequately calibrated. 30 , 31 Additionally, these studies identified equipment in use that had bulb malfunction, tubing tears, and excessive wear, all of which impact the physical therapist’s ability to obtain an accurate BP reading. Regular calibration and inspection of equipment at regular intervals ranging from 6 months to 2 years is essential to assure BP measures are not under- or overestimated.…”
Section: Recommendations For Bp Screening In Outpatient Physical Thermentioning