Abstract: The purpose of this study was to provide an empirical model to develop reference air kerma (RAK) alert levels as a function of patient thickness or age for pediatric fluoroscopy for any institution to use in a Quality Assurance program. RAK and patient thickness were collected for 10&663 general fluoroscopic examinations and 1500 fluoroscopically guided interventions (FGIs). RAK and patient age were collected for 6137 fluoroscopic examinations with mobile-C-arms (MC). Coefficients of linear regression fits of … Show more