The measured accuracies of the dose distributions calculated by the convolution algorithm of the LGP were within the clinically acceptable range (GIPR ≥ 96.9%) for various configurations of collimator size, location, direction of calculation plane, and number of shots. Due to the intrinsic asymmetry in the dose distribution along the z-axis, the treatment plan should also be verified in coronal or sagittal plane.