Artifacts in magnetic force microscopy under in-plane applied magnetic field: Magnetic bubble as a case study

Abstract: The study of magnetic bubble configuration in Co/Ni multilayers circular dots under in-plane magnetic field enlightens two magnetic force microscopy (MFM) measurement artifacts. We demonstrate that any tilt of the magnetic field producing out-of-plane field component due to the inhomogeneity of the MFM set up strongly affects the shape and size of the magnetic bubble independently of the in-plane field component intensity. Furthermore, MFM signal variations for inplane magnetic field larger than 0.1 Tesla can … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
4

Relationship

1
3

Authors

Journals

citations
Cited by 4 publications
(2 citation statements)
references
References 18 publications
0
2
0
Order By: Relevance
“…As an example, the assessment of iron deficiency in the human body typically relies on the measurement of ferritin protein levels [94]. It is important to note that MFM is not artifact-free [95][96][97]. Previous studies on an MFM analysis of biological materials have revealed a significant challenge, namely, the potential contamination of the signal due to artifacts arising from topographical cross-talk.…”
Section: Afm Imaging Artifactsmentioning
confidence: 99%
“…As an example, the assessment of iron deficiency in the human body typically relies on the measurement of ferritin protein levels [94]. It is important to note that MFM is not artifact-free [95][96][97]. Previous studies on an MFM analysis of biological materials have revealed a significant challenge, namely, the potential contamination of the signal due to artifacts arising from topographical cross-talk.…”
Section: Afm Imaging Artifactsmentioning
confidence: 99%
“…In this attempt to increase the sensitivity of magnetic images, the presence of artifacts when measuring non-conventional magnetic samples must be considered [28,29]. Those artifacts can have undesirable effects on magnetic images leading to misinterpretations of the signal [30]. In general, when imaging magnetic materials with MFM the most typical artifact is the crosstalk between the topography and the magnetic signal.…”
Section: Introductionmentioning
confidence: 99%