Assessment of Carbon/Salt/Adhesive Electrodes for Surface Electromyography Measurements

Abstract: This paper presents the evaluation of novel electrodes for surface electromyography (sEMG) measurements. The electrodes are based on the mixture of carbon powder, quaternary salt, and viscoelastic polymeric adhesive (carbon/salt/adhesive or simply CSA), which when combined, provide the unique advantages of having longer (theoretically infinite) shelf life and potentially lower cost than Ag/AgCl hydrogel electrodes, consistent with FLEXcon’s Patent #8 673 184. The 20 subjects were recruited to collect simultane… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
21
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 38 publications
(22 citation statements)
references
References 13 publications
1
21
0
Order By: Relevance
“…SNR for the ESS electrode was greater than 25 dB across all tests, and there was no apparent trend in signal quality with an SNR > 30 dB over four days, and the average CV was below 10% for all muscles types and tests conducted. This is much higher than the suggested 15 dB for sEMG data with noise artifacts, and higher than or comparable to SNR reported in previous literature [ 18 , 24 ]. In fact, the average SNR of both the six-hour trial and the four-day trial were close to 50 dB, the SNR of an sEMG signal under ideal conditions [ 31 ], and the ESS SNR was significantly higher than the Delsys SNR for the ECR, FCU, and TRI during the six-hour test.…”
Section: Discussionsupporting
confidence: 64%
See 2 more Smart Citations
“…SNR for the ESS electrode was greater than 25 dB across all tests, and there was no apparent trend in signal quality with an SNR > 30 dB over four days, and the average CV was below 10% for all muscles types and tests conducted. This is much higher than the suggested 15 dB for sEMG data with noise artifacts, and higher than or comparable to SNR reported in previous literature [ 18 , 24 ]. In fact, the average SNR of both the six-hour trial and the four-day trial were close to 50 dB, the SNR of an sEMG signal under ideal conditions [ 31 ], and the ESS SNR was significantly higher than the Delsys SNR for the ECR, FCU, and TRI during the six-hour test.…”
Section: Discussionsupporting
confidence: 64%
“…To help address these concerns, a variety of dry electrodes have been developed with flexible polymers and textiles that allow the electrode to adhere and move with the skin. For electrodes with a polymer adhesive, thin sheets of metal [ 10 , 18 , 19 , 20 , 21 , 22 , 23 ], carbon ink [ 24 ] and powders [ 25 ], as well as atomically thin sheets such as graphene [ 26 ] have been used as a conductive material. Textile-based electrodes with silver-plated conductive thread have also been reported [ 27 ].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“… Signal-to-motion ratio (SMR) gives information about the level of the motion artifacts in the signal. It was computed using the whole recorded signal according to the description given by [ 42 ] where is the power to the frequency in the power spectral density (PSD) and is a straight line between cero and the highest mean power density, is the sampling frequency: 5 kHz. Zero-crossing (ZC) provides indirect information about the time domain of the signal frequency.…”
Section: Methodsmentioning
confidence: 99%
“…Informed consent was obtained from each subject prior to their participation in the experiment. A wet contact electrode (Ag-AgCl) is the gold standard of a surface EMG measurement 16 . It was used to benchmark the performance of these porous medical bandage insulated cEMG biomedical sensors 17 .…”
Section: Hz Digital Notch Filter Design Parametermentioning
confidence: 99%