Assessment of MRI image distortion based on 6 consecutive years of annual QAs and measurements on 14 MRI scanners used for radiation therapy

Abstract: Purpose To determine the magnitude of MRI image distortion based on 6 consecutive years of annual quality assurances/measurements on 14 MRI scanners used for radiation therapy and to provide evidence for the inclusion of additional margin for treatment planning. Methods and materials We used commercial MRI image phantoms to quantitatively study the MRI image distortion over period of 6 years for up to 14 1.5 and 3 T MRI scanners that could potentially be used to provide… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

4
3
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 6 publications
(7 citation statements)
references
References 22 publications
4
3
0
Order By: Relevance
“…These values can be used as an estimate for the magnitude of system‐related distortions on both the Sola and Aera platforms. The 3D distortion was found to be stable over 2 years in agreement with Lu et al 16 . and should be performed annually and after major upgrades.…”
Section: Discussionsupporting
confidence: 88%
See 4 more Smart Citations
“…These values can be used as an estimate for the magnitude of system‐related distortions on both the Sola and Aera platforms. The 3D distortion was found to be stable over 2 years in agreement with Lu et al 16 . and should be performed annually and after major upgrades.…”
Section: Discussionsupporting
confidence: 88%
“…These values were found to be within the recommendations of report 284 19 and the measured distortion showed a clear downward trend on the Sola compared to the Aera.These values can be used as an estimate for the magnitude of system-related distortions on both the Sola and Aera platforms. The 3D distortion was found to be stable over 2 years in agreement with Lu et al 16 and should be performed annually and after major upgrades. 3D distortion phantoms can be used to estimate and optimize the distortion of clinical sequences, for example, the brain stereotactic radiosurgery sequence was optimized to a maximum distortion of 1 mm.…”
Section: Discussionsupporting
confidence: 87%
See 3 more Smart Citations