Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss

Abstract: SummaryNon-invasive monitoring of haemoglobin concentration provides real-time measurement of haemoglobin concentration (SpHb) using multi-wavelength pulse co-oximetry. We hypothesised that in-vivo adjustment using the mean of three haemoglobinometer (HemoCue â ) measurements from an arterial blood sample at the first SpHb measurement (HCueART) would increase the accuracy of the monitor. The study included 41 adults for a total of 173 measurements of haemoglobin concentration. In-vivo adjusted SpHb was automat… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
12
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 14 publications
(12 citation statements)
references
References 17 publications
0
12
0
Order By: Relevance
“…However, the accuracy of Hcue ven was found to be close to that of Hb(gas); the agreement of results of the Hcue cap method with those of the reference method were distinctly worse. Previous studies among adult patients in the ICU and in the operating theatre have reported differences in the accuracy of HemoCue method that depended on the source of blood samples . To our knowledge, comparison studies of Hcue cap and Hcue ven in neonates have not been conducted.…”
Section: Discussionmentioning
confidence: 92%
“…However, the accuracy of Hcue ven was found to be close to that of Hb(gas); the agreement of results of the Hcue cap method with those of the reference method were distinctly worse. Previous studies among adult patients in the ICU and in the operating theatre have reported differences in the accuracy of HemoCue method that depended on the source of blood samples . To our knowledge, comparison studies of Hcue cap and Hcue ven in neonates have not been conducted.…”
Section: Discussionmentioning
confidence: 92%
“…This is a frequent situation in pediatric anesthesia, because arterial lines or central venous catheters are less common in pediatric patients than in adult patients. The accuracy of capillary Hemocue is not equivalent to venous Hemocue or blood gas analysis, but superior to noninvasive Hb measurement methods . To summarize, noninvasive and blood‐sparing bed‐side Hb measurement methods are rarely applied in German pediatric anesthesia.…”
Section: Discussionmentioning
confidence: 99%
“…The accuracy of capillary Hemocue is not equivalent to venous Hemocue or blood gas analysis, but superior to noninvasive Hb measurement methods. 35,36 To summarize, noninvasive and blood-sparing bed-side Hb measurement methods are rarely applied in German pediatric anesthesia.…”
Section: Discussionmentioning
confidence: 99%
“…The accuracy and reliability of the portable Pronto-7 Pulse COOximeter was reported in numerous studies conducted in the operating room [10][11][12][13][14][15][16][17][18][19][20][21] and in the intensive care unit [4,5,[22][23][24][25] with predominantly satisfactory performance. In a study done in bleeding trauma patients the agreement was moderate [bias (LOA) of -0.6 (-3.0; +1.9)] g/dl [26], where the limit of agreement is defined as the mean difference + twice the standard deviation of the differences [27].…”