AIM: To evaluate the diagnostic reliability of a new generation wireless point-of care ultrasound device for abdominal and thoracic findings. MATERIAL AND METHODS: 40 patients (16 females, 24 males 19 –80 years, on average 56.1 years) were scanned by an experienced examiner using the new wireless Vscan Air device for abdominal and thoracic findings. The probe frequencies were 2–5 MHz (convex probe) and 3–12 MHz for the linear probe. As a reference standard, all patients were also examined using high-end ultrasound (LOGIQ E9/10). Results were interpreted independently by two examiners in consensus, also with regard to the image quality (0–4, from not assessable = 0, to excellent 4). RESULTS: In all 40 patients (100%) examination with conventional high-end ultrasound and the Vscan Air ultrasound device was feasible. Sensitivity, specificity, positive and negative predictive value for the diagnosis of abdominal and thoracic findings were 63.3%, 100%, 100%, and 40%, respectively. Most main diagnostic findings were detected using the mobile device compared to the high-end ultrasound. Limitations were found regarding characterization and classification of hepatic and renal tumorous lesions. Image quality revealed mostly minor diagnostic limitations for the mobile device, mean 2.9 (SD –0.300) and was excellent or with only minor diagnostic limitations for conventional high-end ultrasound, mean 3.25 (SD –0.438). CONCLUSION: Due to its easy application and its high diagnostic reliability, point-of-care ultrasound systems of the latest generation represent a valuable imaging method for the primary assessment of abdominal and thoracic findings, especially in patients on intensive care units or in emergency situations.