CT image quality over time: comparison of image quality for six different CT scanners over a six‐year period

Abstract: UNSCEAR concluded that increased use of CT scanning caused dramatic changes in population dose. Therefore, international radiation protection authorities demand: 1) periodical quality assurance tests with respect to image quality and radiation dose, and 2) optimization of all examination protocols with respect to image quality and radiation dose. This study aimed to evaluate and analyze multiple image quality parameters and variability measured throughout time for six different CT scanners from four different … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
17
0

Year Published

2017
2017
2023
2023

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 33 publications
(17 citation statements)
references
References 12 publications
0
17
0
Order By: Relevance
“…Site-specific CT-to-density calibrations are often required due to variations between CT scanners, techniques and reconstruction algorithms. Different CT scanner models and manufacturers have been shown to produce different measurements [1] , [2] , [3] , [4] , [5] , [6] , [7] , [8] , [9] . For a given scanner, choice of tube voltage can have a significant impact on measured CT numbers [1] , [2] , [3] , [4] , [5] , [8] , [10] , [11] , [12] while other parameters such as tube current have been shown to have minimal impact [4] , [12] .…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Site-specific CT-to-density calibrations are often required due to variations between CT scanners, techniques and reconstruction algorithms. Different CT scanner models and manufacturers have been shown to produce different measurements [1] , [2] , [3] , [4] , [5] , [6] , [7] , [8] , [9] . For a given scanner, choice of tube voltage can have a significant impact on measured CT numbers [1] , [2] , [3] , [4] , [5] , [8] , [10] , [11] , [12] while other parameters such as tube current have been shown to have minimal impact [4] , [12] .…”
Section: Introductionmentioning
confidence: 99%
“…Beyond scanner parameters, variables including software [13] , reconstruction algorithm [3] , [14] and object location [15] have also been demonstrated to alter CT numbers. Once a CT scanner has been calibrated properly, stable CT number over time can be achieved [6] and variations due to phantom variation are typically much lower than those from using different scanners and tube voltages [16] .…”
Section: Introductionmentioning
confidence: 99%
“…Therefore, for various contrast objects (teflon, delrin, LDPE, polystyrene, acrilic), we assessed CNR as the ratio between contrast (i.e. difference in CT numbers between the object and background) and noise [ 51 , 67 , 68 , 79 ]. Furthermore, we included the LCD analysis, which allows to assess the limiting detectable contrast threshold for a given object size [ 71 , 80 83 ].…”
Section: Discussionmentioning
confidence: 99%
“…Increased image noise impeded clinical interpretation of post‐contrast CT scans from an updated CT scanner that was recently added to a fleet of existing scanners from the same vendor. The goals of this work were to quantify and standardize image quality on the new and an existing scanner using phantom images and determine why this issue was not discovered upon acceptance testing, which used conventional techniques employing water or water‐equivalent uniform sections of phantom to quantify image noise …”
Section: Discussionmentioning
confidence: 99%
“…These advancements create challenges in harmonizing image quality and appearance from a large fleet of clinical scanners. Several groups, however, have quantified and homogenized image appearance on similar scanner models using uniform sections of water or water‐equivalent materials . At our institution, feedback from radiologists indicated that differences in image appearance and noise on a new CT scanner we recently added to our fleet impeded reading of post‐contrast scans (Fig.…”
Section: Introductionmentioning
confidence: 99%