Spatial aperture coding is a technique used to improve
throughput
without sacrificing resolution both in optical spectroscopy and sector
mass spectrometry (MS). Previous work demonstrated that aperture coding
combined with a position-sensitive array detector in a miniature cycloidal
mass spectrometer was successful in providing high-throughput, high-resolution
measurements. However, due to poor alignment and field nonuniformities,
reconstruction artifacts were present. Recently, significant progress
was made in eliminating most of the reconstruction artifacts with
improved field uniformity and alignment. However, artifacts as large
as 1/3 of the main peak were still observed at low mass (<17 u).
Such artifacts will reduce accuracy in identification and quantification
of analytes, reducing the impact of the throughput advantage gained
by using a coded aperture. The artifacts were hypothesized to be a
result of a mass dependent in curvature of ions in the ion source.
Ions with higher mass (m/z >
17
u) and a larger curvature did not pass through all slits in the coded
aperture. Therefore, when reconstructing with a system response derived
from the aperture image from a higher mass m/z = 32 u ion, reconstruction artifacts appeared for m/z < 17 u. In this work, two methods
were implemented to significantly reduce the presence of artifacts
in reconstructed data. First, we modified the reconstruction algorithm
to incorporate a mass-dependent system response function across the
mass range (10–110 u). This method reduced the size of the
artifacts by 82%. Second, to validate the hypothesis that the mass-dependent
system response function was a result of differences in curvature
of ions in the ion source, we modified the design of the ion source
by shifting the coded aperture slits relative to the center of the
ionization volume. This method resulted in ions of all masses passing
through all slits in the coded aperture, a constant system response
function across the entire mass range. Artifacts were reduced by 94%.