Difficult Retrieval of a Recovery IVC Filter

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
13
0

Year Published

2008
2008
2016
2016

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 33 publications
(14 citation statements)
references
References 3 publications
1
13
0
Order By: Relevance
“…Cardiovasc Diagn Ther 2016;6(6):642-650 cdt.amegroups.com loops such as the duodenum, aortic wall penetration, and small-bowel volvulus. Filter migration and spontaneous fracture and embolization of filter components have also been reported (8,10,14,27,28). Nonetheless, before using advanced techniques for IVC filter retrieval, the associated possible complications should always be compared to the potential risks of leaving the filter permanently indwelling, and should be individualized.…”
Section: Complications Of Ivc Filter Retrievalsmentioning
confidence: 99%
See 2 more Smart Citations
“…Cardiovasc Diagn Ther 2016;6(6):642-650 cdt.amegroups.com loops such as the duodenum, aortic wall penetration, and small-bowel volvulus. Filter migration and spontaneous fracture and embolization of filter components have also been reported (8,10,14,27,28). Nonetheless, before using advanced techniques for IVC filter retrieval, the associated possible complications should always be compared to the potential risks of leaving the filter permanently indwelling, and should be individualized.…”
Section: Complications Of Ivc Filter Retrievalsmentioning
confidence: 99%
“…Stavropoulos et al later concluded that this endobronchial forceps technique could be safely used to successfully remove embedded filters (21,22). In this technique a rigid endobronchial forceps is introduced through a large caliber venous access sheath (14)(15)(16)(17)(18) and is used to dissect the hook of the IVC filter free from the caval wall. The tip or hook of the filter may then be grasped with the forceps so that the access sheath can be advanced over the filter apex in order to collapse the filter and allow subsequent removal using the forceps (Figure 8).…”
Section: Dissection Techniques Endobronchial Forceps Dissection and Rmentioning
confidence: 99%
See 1 more Smart Citation
“…Filter fractures are less frequent and are reported less than 1% in the literature 6 . A few studies and case reports have been published about prolonged temporary implantation with use of "Recovery IVC Filter (TM) " [1][2][3][4][5]8,9 . The first report about the use of "Recovery IVC Filter (TM) " was from Asch in the year 2002 and he reported an average period of implantation 53 days (5-134 days) in which out of 32 implanted filters, 26 were successfully retrieved.…”
Section: Discussionmentioning
confidence: 99%
“…The wire is placed adjacent to the tilted filter and used to deflect the apex of the filter into the lumen of the IVC. Conventional snare or cone retrieval commences thereafter [24,70].…”
Section: Loop Snare With Single Accessmentioning