Doppler ultrasound triggering for cardiac MRI at 7T

Abstract: Doppler ultrasound was applied as a new trigger method in CMR at 7T. The transmission line and transducer were locally approved as 7T MR conditional, and were successfully tested for image synchronization at 7T. In the future, this method needs to be evaluated in a larger patient population. Magn Reson Med 80:239-247, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
12
0

Year Published

2017
2017
2023
2023

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 10 publications
(12 citation statements)
references
References 45 publications
(61 reference statements)
0
12
0
Order By: Relevance
“…The Doppler ultrasound (DUS) device and signal processing algorithms used in this work consist of a custom-built DUS unit and a signal processing unit [15,25,26] (north medical GmbH, Hamburg, Germany). The DUS unit transmits 1 MHz pulses with a repetition frequency of 3.2 kHz to a single 1 × 1 cm-piezoelectric element transducer, made from non-magnetic lead zirconate titanate, placed on the subjects' chest.…”
Section: Doppler Ultrasound For Cardiac Synchronizationmentioning
confidence: 99%
See 4 more Smart Citations
“…The Doppler ultrasound (DUS) device and signal processing algorithms used in this work consist of a custom-built DUS unit and a signal processing unit [15,25,26] (north medical GmbH, Hamburg, Germany). The DUS unit transmits 1 MHz pulses with a repetition frequency of 3.2 kHz to a single 1 × 1 cm-piezoelectric element transducer, made from non-magnetic lead zirconate titanate, placed on the subjects' chest.…”
Section: Doppler Ultrasound For Cardiac Synchronizationmentioning
confidence: 99%
“…In addition, a resonant circuit is placed at the end of the transmission line. A safety assessment was performed in [25] for the case where the transducer was placed within the transmitting volume of a 7T cardiac array coil. In the present work, the transducer is placed more than 25 cm from the end of the transmitting head coil, where the E-and H-fields are significantly reduced compared to the placement in [25].…”
Section: Doppler Ultrasound For Cardiac Synchronizationmentioning
confidence: 99%
See 3 more Smart Citations