We studied the full-energy peak efficiency of a Si drift detector (SDD) and a Si (Li) detector (SLD) using the formalisms proposed by Seltzer [Nucl. Instrum. Meth. 188 (1981) 133] and O'Meara and Campbell [X-Ray Spectrom. 33 (2004) 146]. The respective adjustable parameters were fitted to full-energy peak efficiencies measured with certified point radioactive sources. Seltzer's model was able to reproduce the experimental data for the SDD and the SLD in the 6-40 keV and 6-100 keV energy ranges, respectively. In turn, O'Meara and Campbell's formula also performed well for the SDD between 6 and 40 keV and proved to fit satisfactorily in the energy interval 6-60 keV for the SLD.