1984
DOI: 10.1017/s0195941700060173
|View full text |Cite
|
Sign up to set email alerts
|

Endotoxin Contamination of Foley Catheters Associated With Fevers Following Transurethral Resection of the Prostate

Abstract: In 1980 an investigation was conducted due to an apparent cluster of fevers following transurethral resections of the prostate (TURP), where the presence of prostatitis or urinary tract infections (UTIs) could not account for most cases, and the usage of antimicrobials did not prevent them. When unused, prepackaged, sterilized samples from 13 lots of triple lumen catheters were analyzed, four were found to contain high levels of endotoxin (ET), with a range of 49,150-≥ 6.25 ng/catheter. Additionally, these lot… Show more