Evaluation of the OC-SENSOR neo System for Testing Fecal Occult Blood

Abstract: Background : Fecal occult blood tests (FOBTs) have been widely used as a means of colorectal cancer screening. Automated FOBTs using immunologic principles have the advantages such as quantitation, high specificity, and high throughput. We evaluated a newly-introduced automated FOBT analyzer, OC-SENSOR neo (OC neo) (Eiken Chemical Co., Japan).Methods : The precision, linearity, and carry-over rate of OC neo were assessed with specimens prepared in accordance with the guidelines of CLSI. We performed a parallel… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
3
0
1

Year Published

2009
2009
2021
2021

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(5 citation statements)
references
References 15 publications
1
3
0
1
Order By: Relevance
“…However, both products showed a slightly higher % difference, 3.0∼6.0% in the middle range of AMR 50 to 1000 ng/mL. These results were similar to those in previous studies in which OC sensor series were evaluated, and it was thought that there was no significant difference in linearity between LX-CS and OC-CS [22].…”
Section: Linearity Analysis Of Lx-cs and Oc-cs Measuredsupporting
confidence: 90%
See 1 more Smart Citation
“…However, both products showed a slightly higher % difference, 3.0∼6.0% in the middle range of AMR 50 to 1000 ng/mL. These results were similar to those in previous studies in which OC sensor series were evaluated, and it was thought that there was no significant difference in linearity between LX-CS and OC-CS [22].…”
Section: Linearity Analysis Of Lx-cs and Oc-cs Measuredsupporting
confidence: 90%
“…In a previous study that evaluated OC NEO and OC DIANA among the OC sensor series, it can be seen that the precision was higher than that measured at 7.8% and 3.1% for low-concentration materials, and 3.5% and 2.0% for high-concentration materials, respectively [19]. Besides, in a previous study evaluating HM-JACK (Kyowa Medex, Japan) and OC IO (Eiken Chemical, Japan) at various concentrations, repeatability of positive specimens using HM JACK is 3.5% in high concentration materials and 4.8% in low concentration specimens [20].…”
Section: Discussionmentioning
confidence: 86%
“…2), in the Department of Colorectal Surgery, Singapore General Hospital. This machine, which employs the same technology as the OC-SENSOR MICRO used in several other studies [10,13,16], has been validated in a Korean study [14] and has been registered for use internationally since April 2008. Based on the manufacturer's recommendations, a faecal occult blood level of C100 ng Hb/ml was taken as a positive result.…”
Section: Study Population and Designmentioning
confidence: 99%
“…The iFOBT has been evaluated in Japan and elsewhere [12][13][14][15] and has replaced the guaic-based FOBT in several screening programmes worldwide [12][13][14][15][16]. Furthermore, automated quantitative immunochemical FOBT (qFOBT) allows for the selection of a threshold for colonoscopy, making it an excellent screening tool.…”
Section: Introductionmentioning
confidence: 98%
“…또한 guaiac법에 비해서 높은 민감도와 특이도를 보이며 [8], 정량이 가능해서 임계치를 설 정하여 양성 판정을 객관적으로 할 수 있다. 국내에서는 최근 임상 검사에서 10여 종의 자동화된 정량 분변잠혈검사가 도입되어 사용 중이나 [6], 성능평가는 부족하며 OC-Sensor 시리즈(Eiken Chemical, Tokyo, Japan) 중 NEO (Eiken Chemical)에 대한 성능평가 논 문만 있다 [9].…”
unclassified