A simple and rapid high-performance liquid chromatography with diode array detector (HPLC-DAD) method has been developed and validated for simultaneous quantification of five antiemetic agents in infusion samples: dexamethasone, ondansetron, granisetron, tropisetron, and azasetron. The chromatographic separation was achieved on a Phenomenex C18 column (4.6 mm × 150 mm, 5 μm) using acetonitrile-50 mM KH2PO4 buffer-triethylamine (25 : 74 : 1; v/v; pH 4.0). Flow rate was 1.0 mL/min with a column temperature of 30°C. Validation of the method was made in terms of specificity, linearity, accuracy, and intra- and interday precision, as well as quantification and detection limits. The developed method can be used in the laboratory to routinely quantify dexamethasone, ondansetron, granisetron, tropisetron, and azasetron simultaneously and to evaluate the physicochemical stability of referred drugs in mixtures for endovenous use.