Imaging Parameter Optimization of 3D Phase Contrast-MRA to Reduce Susceptibility-artifact and Radiofrequency-shielding around the Intracranial Stent

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2020
2020
2021
2021

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(7 citation statements)
references
References 15 publications
0
7
0
Order By: Relevance
“…24) The disadvantages of SILENT MRA were as follows: the total required time is longer in SILENT MRA than that in TOF-MRA because SILENT MRA needs ASL preparation 7) and subtraction processing. 17) As another problem, the signal in UTE-MRA is weaker than that in TOF-MRA, especially in the distal part from the neck where blood was labeled. 3) Furthermore, ASL may not always be performed if there is severe neck carotid artery stenosis or metal around the neck, such as an intubation tube cuff.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…24) The disadvantages of SILENT MRA were as follows: the total required time is longer in SILENT MRA than that in TOF-MRA because SILENT MRA needs ASL preparation 7) and subtraction processing. 17) As another problem, the signal in UTE-MRA is weaker than that in TOF-MRA, especially in the distal part from the neck where blood was labeled. 3) Furthermore, ASL may not always be performed if there is severe neck carotid artery stenosis or metal around the neck, such as an intubation tube cuff.…”
Section: Discussionmentioning
confidence: 99%
“… 12) Also, the cost and the availability of the 3 tesla magnetic resonance imaging (MRI) are problems. 17) Furthermore, in this study, all patients were treated using the fourth-generation YASARGIL Aneurysm Clip System made of TiAl6V4 titanium alloy (ISO 5832-2). We could not investigate metal artifacts from other clips made of other materials.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…However, the total required time is longer in UTE-MRA than that in TOF-MRA because UTE-MRA needs arterial spin labeling (ASL) preparation,[ 8 ] multiple scans, and subtraction processing. [ 6 ] As another problem, the signal in UTE-MRA is weaker than that in TOF- MRA. [ 3 ] Furthermore, ASL may not always be performed if there is metal around the neck, such as an intubation tube cuff.…”
Section: Discussionmentioning
confidence: 99%
“…Compared to 1.5T UTE-MRA, 3.0T UTE- MRA has the better image quality and wider description range, but the cost and the availability of the 3.0T MRI are problems. [ 6 ] In summary, 3.0T UTE-MRA can reduce the time and the invasiveness compared to CTA as well as overcome the disadvantages of 1.5T UTE-MRA, but it also has some disadvantages compared to the other sequences. In addition, the clip-induced metal artifact should be further investigated in 3.0T UTE-MRI,[ 11 ] and it should be clarified what kind of aneurysm, clip shape, and surgical procedure, like how to apply the clip, can be applied to 3.0T UTE-MRI.…”