Immunoassay Interference by a Commonly Used Blood Collection Tube Additive, the Organosilicone Surfactant Silwet L-720

Abstract: Background: A small number of immunoassays on several different types of analyzers were recently adversely affected by tube additives in Becton Dickinson (BD) Vacutainer® SST™, SST II, and Microtainer™ blood collection tubes. We examined the effect of a commonly used tube surfactant, Silwet™ L-720, on immunoassays and the mechanism for the interference. Methods: Immunoassays were performed on serum supplemented with Silwet L-720 on the IMMULITE™ 2500 and AxSYM™ analyzers. Direct effects of the s… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

4
46
0

Year Published

2005
2005
2024
2024

Publication Types

Select...
5
3
1

Relationship

0
9

Authors

Journals

citations
Cited by 52 publications
(50 citation statements)
references
References 26 publications
4
46
0
Order By: Relevance
“…18 More recently, the same investigators identi¢ed the source of the interference with the IMMULITEt 2500 TT 3 assay to be the high concentration of surfactant coated on the inner surface of the tubes, and suggested desorption of solid phase antibodies employed in the assays as the interfering mechanism. 7 In competitive solid-phase immunoassays, where antigen in the sample competes with labelled antigen for binding to solid-phase antibody, release of antibodies from the solid phase would result in a decrease in the signal which would be interpreted as a higher concentration of antigen being present. Except for CA15-3, for all immunoassays that were cited by the MHRA to give a positive bias, Advia Centaur s employs competitive solid-phase immunoassays.…”
Section: Discussionmentioning
confidence: 99%
“…18 More recently, the same investigators identi¢ed the source of the interference with the IMMULITEt 2500 TT 3 assay to be the high concentration of surfactant coated on the inner surface of the tubes, and suggested desorption of solid phase antibodies employed in the assays as the interfering mechanism. 7 In competitive solid-phase immunoassays, where antigen in the sample competes with labelled antigen for binding to solid-phase antibody, release of antibodies from the solid phase would result in a decrease in the signal which would be interpreted as a higher concentration of antigen being present. Except for CA15-3, for all immunoassays that were cited by the MHRA to give a positive bias, Advia Centaur s employs competitive solid-phase immunoassays.…”
Section: Discussionmentioning
confidence: 99%
“…In response to the immunoassay interference, the manufacturer has reformulated the blood collection tubes to reduce the amount of tube additives [31] .…”
Section: Blood Collection Tube Interference On Clinical Chemistry Assaysmentioning
confidence: 99%
“…One of the interferents for immunoassays has been identified as an organosilicone surfactant (5 ). Daae et al (6,7 ) investigated the difference of hematology values in capillary and venous blood samples collected in Microtainers and Vacutainers, respectively.…”
Section: To the Editormentioning