Impact of storage temperature and time before analysis on electrolytes (Na+, K+, Ca2+), lactate, glucose, blood gases (pH, pO2, pCO2), tHb, O2Hb, COHb and MetHb results
Abstract:Objectives
The objective of our study is to evaluate the effect of storage temperature and time to analysis on arterial blood gas parameters in order to extend the CLSI recommendations.
Methods
Stability of 12 parameters (pH, pCO₂, pO₂, Na+, K+, Ca2+, glucose, lactate, hemoglobin, oxyhemoglobin, carboxyhemoglobin, methemoglobin) measured by GEM PREMIER™ 5000 blood gas analyzer was studied at room temperature and at +4 °C (52 … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.