“…Salbutamol entrained on the filter was described as total inhaled dose (aerosol emitted) and assayed for salbutamol using previously validated highperformance liquid chromatography. 12,13,25,26,32,33 We used the 25 mm ϫ 4.6 mm Zorbax Eclipse Plus C18 and the ODS1 column (Agilant), through which the mobile phase consisted of a mixture of acetonitrile and water (0.1% phosphoric acid) (90:10, v/v) was pumped at 1 mL/min using the 1260 Infinity preparative pump (G1361A, Agilant) the 1260 Infinity Diode array detector VL (G131SD, Agilant) was set at 225 nm with an injection volume 100 L. Calibration solutions ranged from 4 to 100 g/mL (weight/volume). Limit of detection was 0.35 g/mL, and the lower limit of quantification was 2.55 g/mL.…”