High-throughput liquid chromatography with tandem mass spectrometric detection (LC-MS/MS) methodology for the determination of methamphetamine (METH), amphetamine (AMP), 4-hydroxymethamphetamine (4-OH-METH), and 4-hydroxyamphetamine (4-OH-AMP) was developed and validated using simple trichloroacetic acid sample treatment. The method was validated in rat serum, brain, and testis. Lower limits-of-quantitation (LOQ) for METH and AMP were 1 ng • mL − 1 using positive ion electrospray tandem mass spectrometry (MS/MS). The accuracy of the method was within 25% of the actual values over a wide range of analyte concentrations. The within-assay precision was better than 12% (co effi cient of variation). The method was linear over a wide dynamic range (0.3-1000 ng • mL − 1 ). Quantitation was possible in all 3 matrices using only serum standards because of minimal matrix-associated ion effects or the use of an internal standard. Finally, the LC-MS/MS method was used to determine serum, brain, and testis METH and AMP concentrations during a subcutaneous infusion (5.6 mg kg − 1 day − 1 ) of METH in rats. Concentrations of 4-OH-AMP and 4-OH-METH were below the LOQ in experimental samples. The bias introduced by using serum calibrators for the determination of METH and AMP concentrations in testis and brain was less than 8% and insignifi cant relative to the interanimal variability.K EYWORDS: methamphetamine , rats , LC-MS/MS , matrix ion effects