2009
DOI: 10.1016/j.jvs.2009.04.026
|View full text |Cite
|
Sign up to set email alerts
|

Long-term results of balloon-expandable LifePath endografts in abdominal aortic aneurysm: A single-center experience

Abstract: Wire form fracture is the major structural problem in the LifePath balloon-expandable endograft device, resulting in a significant endoleak rate. We must caution those patients with a LifePath device in-situ that careful follow-up must be performed due to material fatigue and they should consider secondary conversion.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
1
0

Year Published

2009
2009
2010
2010

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 16 publications
0
1
0
Order By: Relevance
“…20 The long-term fate of an aortic neck treated with a balloon-expandable stent-graft requires further investigation, 2124 and the properties of this type of device may be beneficial, especially for difficult aneurysms with complex anatomy. 2528 The balloon-expandable design offers the advantages of precise positioning and variable radial force, 2935 so balloon expansion may be a better fixation method. We report the clinical results of our first 30 patients treated with a balloon-expandable stent-graft, with particular attention to the evolution of the proximal aortic neck during a follow-up ranging from 4 to 8 years.…”
mentioning
confidence: 99%
“…20 The long-term fate of an aortic neck treated with a balloon-expandable stent-graft requires further investigation, 2124 and the properties of this type of device may be beneficial, especially for difficult aneurysms with complex anatomy. 2528 The balloon-expandable design offers the advantages of precise positioning and variable radial force, 2935 so balloon expansion may be a better fixation method. We report the clinical results of our first 30 patients treated with a balloon-expandable stent-graft, with particular attention to the evolution of the proximal aortic neck during a follow-up ranging from 4 to 8 years.…”
mentioning
confidence: 99%