On the response of a europium doped phosphor-coated CMOS digital imaging detector

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
26
1

Year Published

2014
2014
2019
2019

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 36 publications
(27 citation statements)
references
References 50 publications
0
26
1
Order By: Relevance
“…This value is higher than what has been reported in literature that is 10% MTF at 6 cycles/mm for a CMOS/Gd 2 O 2 S:Eu (65.1 mg/cm 2 ) combination [3]. However other CMOS/scintillator combinations such as CMOS/Gd 2 O 2 S:Eu (33.3 mg/cm 2 ) and CMOS/Lu 2 O 3 :Eu (65.1 mg/cm 2 ) demonstrates 10% MTF above 11 cycles/mm [4].The exposure conditions in these cases were not exactly the same since the 70kVp X-ray spectrum was hardened by the addition of 21 mm Al [3,4] Other dental detectors, by means of viewing the image of a bar pattern, have reported a resolution limit above 10 cycles/mm [8]. In figure 3 the NNPS curves at 60 kVp, 104.8 µGy and 70 kVp, 100.8 µGy are demonstrated.…”
Section: Resultscontrasting
confidence: 64%
See 1 more Smart Citation
“…This value is higher than what has been reported in literature that is 10% MTF at 6 cycles/mm for a CMOS/Gd 2 O 2 S:Eu (65.1 mg/cm 2 ) combination [3]. However other CMOS/scintillator combinations such as CMOS/Gd 2 O 2 S:Eu (33.3 mg/cm 2 ) and CMOS/Lu 2 O 3 :Eu (65.1 mg/cm 2 ) demonstrates 10% MTF above 11 cycles/mm [4].The exposure conditions in these cases were not exactly the same since the 70kVp X-ray spectrum was hardened by the addition of 21 mm Al [3,4] Other dental detectors, by means of viewing the image of a bar pattern, have reported a resolution limit above 10 cycles/mm [8]. In figure 3 the NNPS curves at 60 kVp, 104.8 µGy and 70 kVp, 100.8 µGy are demonstrated.…”
Section: Resultscontrasting
confidence: 64%
“…PVR characterizes the linearity of the detector response with respect to X-ray exposure. In addition spatial frequency image quality metrics are available like MTF, demonstrating the ability of the detector to resolve objects of different dimensions, as well as, NNPS which demonstrates the relative noise components in spatial frequency domain [3,4]. In current work a digital detector SCHICK CDR CMOS was examined.…”
Section: Introductionmentioning
confidence: 99%
“…For all the ROIs taken from each image, 2D fast Fourier transform (FFT) of each ROI was calculated and added to the NPS ensemble. NNPS was obtained by dividing NPS by the square of the corresponding mean pixel value, and afterward, the ensemble average was obtained [21,22].…”
Section: Normalized Noise Power Spectra (Nnps)mentioning
confidence: 99%
“…This scintillator was selected due to its efficiency and imaging properties, which are superior in comparison to other detectors [17]. The density of Gd 2 O 2 S:Tb is 7.34 g/cm 3 .…”
Section: Methodsmentioning
confidence: 99%