Abstract:This study aims to optimize the iterative deconvolution correction of air-filled
ionization chamber measurements with limited spatial resolution for photon beam dosimetry. The
ionization chamber volume effect could be explained by the inhomogeneity of the absorbed dose
in its sensitive volume, leading to a volume averaging effect acting as a convolution kernel
K(x). Therefore, the “true” dose profile Pt(x) can be deduced by deconvolving measured
1D profiles Pm(x). The Semiflex 3D ionization chambe… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.