Patient and personnel dosimetry in endovascular radiotherapy with 90Sr/90Y sources

Abstract: Endovascular brachytherapy (EVBT) is an established treatment to reduce the probability of restenosis after a percutaneous coronary intervention. The purpose of this study was to assess (1) the manufacturer's stated dosimetric data for (90)Sr/(90)Y source trains to be used in EVBT and (2) the procedure-related radiation burden. The radiation fields in water around six (90)Sr/(90)Y source trains were studied using phantoms made of 'solid water' and MD55-2 radiochromic films. The water equivalence of the phantom… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
5
0

Year Published

2008
2008
2013
2013

Publication Types

Select...
4

Relationship

2
2

Authors

Journals

citations
Cited by 4 publications
(5 citation statements)
references
References 24 publications
0
5
0
Order By: Relevance
“…Several studies[2633] have been performed for the verification of the dose rates of the 90 Sr/ 90 Y source trains, produced by the Novoste Corporation, using radiochromic film dosimetry [Table 2]. One of them is the study of Duggan et al,[26] where four 90 Sr/ 90 Y source trains (5.0 F-30 mm) and an HD-810–type radiochromic film were used.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Several studies[2633] have been performed for the verification of the dose rates of the 90 Sr/ 90 Y source trains, produced by the Novoste Corporation, using radiochromic film dosimetry [Table 2]. One of them is the study of Duggan et al,[26] where four 90 Sr/ 90 Y source trains (5.0 F-30 mm) and an HD-810–type radiochromic film were used.…”
Section: Discussionmentioning
confidence: 99%
“…Most recently, Kalef-Ezra et al . [33] have studied the dosimetry of five 90 Sr/ 90 Y source trains (3.5 F–40-60 mm and 5.0 F–40-60 mm) using MD55-2–type radiochromic film, and they reported that the average of the measured dose rate was 8% higher than the manufacture's stated value. Although the results of these studies are close to certificate values, major differences (exceeding 10%) have also been reported in the study by Piessens and Reynaert.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…These measurements were combined with long-term environmental and personnel dose measurements. Among others, personnel badges containing LiF:Mg,Ti,… thermoluminescent dosimeters (TLDs) measured by either the Greek Atomic Energy Commission (GAEC) or the University of Ioannina Medical Physics Laboratory (UIMPL) [8] were used to monitor occupational and environmental radiological burden during the study period. The adequacy of the room shielding was tested my multiple dose rate measurements carried out at various locations around the treatment using various β, γ and neutron detectors.…”
Section: Methodsmentioning
confidence: 99%
“…According to the standard protocol applied in our laboratory, 6,8 the films are read four days postirradiation. To investigate the long-term stability, films were irradiated at doses between 7 and 20 Gy, kept at room temperature, and read twice, 4 and 30 days postirradiation using either the transmission ͑60 films͒ or the reflection ͑40 films͒ mode.…”
Section: Iid Delay In Readingmentioning
confidence: 99%