Background-Although surgical annuloplasty is the standard repair for ischemic mitral regurgitation (IMR), its application is limited by high morbidity and mortality. Using 2D and real-time 3D echocardiography in an ovine model of chronic IMR, we evaluated the geometric impact and short-term efficacy of a percutaneous transvenous catheter-based approach for mitral valve (MV) repair using a novel annuloplasty device placed in the coronary sinus. Methods and Results-Six sheep developed IMR 8 weeks after induced posterior myocardial infarction. An annuloplasty device optimized to reduce anterior-posterior (A-P) mitral annular dimension and MR was placed percutaneously in the coronary sinus. Mitral annular A-P and commissure-commissure dimensions and MV tenting area (MVTa) in 3 parallel A-P planes (medial, central, and lateral) were assessed by real-time 3D echocardiography with 3D software. The annuloplasty device reduced MR jet area from 5.4Ϯ2.6 to 1.3Ϯ0.9 cm 2 (PϽ0.01), mitral annular A-P dimension in both systole and diastole (24.3Ϯ2.