Performance Evaluation of the FLEX Triumph X-PET Scanner Using the National Electrical Manufacturers Association NU-4 Standards

Abstract: The purpose of this work was to evaluate the performance characteristics of the preclinical X-PET subsystem of the FLEX Triumph PET/CT scanner based on the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). Methods: The performance parameters evaluated include the spatial resolution, scatter fraction, count losses and random coincidences, sensitivity, and image-quality characteristics. The PET detector array consisted of 11,520 individual bismuth germanate crystals arranged in 48 … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

4
49
0

Year Published

2011
2011
2021
2021

Publication Types

Select...
7
1

Relationship

3
5

Authors

Journals

citations
Cited by 44 publications
(53 citation statements)
references
References 32 publications
4
49
0
Order By: Relevance
“…This effect could possibly be attributed to an overshoot at the edges, commonly referred to as the Gibbs effect (29,32). The measured activities in cold regions, that is, the water-and air-filled inserts, for enhanced-resolution mode without postfiltering processing were comparable to data obtained with another scanner (28), suggesting that corrections for attenuation and scatter were done appropriately. In general, preclinical PET attenuation correction is done with CT or with a 57 Co point source by singles detection.…”
Section: Discussionsupporting
confidence: 62%
See 2 more Smart Citations
“…This effect could possibly be attributed to an overshoot at the edges, commonly referred to as the Gibbs effect (29,32). The measured activities in cold regions, that is, the water-and air-filled inserts, for enhanced-resolution mode without postfiltering processing were comparable to data obtained with another scanner (28), suggesting that corrections for attenuation and scatter were done appropriately. In general, preclinical PET attenuation correction is done with CT or with a 57 Co point source by singles detection.…”
Section: Discussionsupporting
confidence: 62%
“…This was somewhat higher than the scatter fractions of currently available preclinical PET scanners (3,28,29), possibly because of the large solid angle of this dedicated breast PET camera and the low-energy threshold for the third and the fourth crystal layers. The energy window of the third and fourth layers was set to 100-800 keV for maintenance of sensitivity.…”
Section: Discussionmentioning
confidence: 83%
See 1 more Smart Citation
“…These methods are widely used and validated on clinical PET systems [11]; however, they still need to be thoroughly investigated on dedicated high resolution, small field-of-view scanners such as those used for small-animal imaging. The quantitative capabilities of the commercially available FLEX Triumph™ (Gamma Medica-Ideas, Nortridge, CA) preclinical PET/CT platform installed in our department are being characterized [12]; however, the lack of reliable image correction methodologies necessitated the implementation of CTAC on this system as a first step to achieve our endeavors. In this work, we implemented and evaluated tube voltage-dependent attenuation correction on this preclinical PET/CT scanner.…”
mentioning
confidence: 99%
“…A large number of small-animal Single Photon Emission Computed Tomography (SPECT) [1][2][3][4][5] and Positron Emission Tomography (PET) [6][7][8][9][10][11][12][13][14][15] systems have been developed and have become commercially available during the past years. Moreover, multimodality imaging has emerged as an important tool in Nuclear Medicine both for diagnostic and therapy follow up imaging.…”
Section: Iintroductionmentioning
confidence: 99%