2015
DOI: 10.1055/s-0035-1555014
|View full text |Cite
|
Sign up to set email alerts
|

Polyethylene Tibial Post Fracture in a Posterior Stabilized (Genesis II) Total Knee Arthroplasty

Abstract: ).Posterior stabilizing knee prostheses have been increasing in popularity recently due to some studies showing excellent long-term survivorship, more consistent kinematics, technical ease, and higher range of motion.1,2 The posterior stabilized knee relies on a cam and post mechanism to replicate native femoral rollback. One of the major potential complications of this system is post fracture. This occurs in less than 1% of most total knees. 3 These patients usually present with a history of a "pop," effusion… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

1
1
0

Year Published

2019
2019
2021
2021

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 9 publications
1
1
0
Order By: Relevance
“…On the post's intraoperative evaluation, there were no signs of significant wear due to post impingement. Similar to other post failure reports in deep flexion, the fracture likely occurred due to increased force on the post during squat exercises [ 23 ]. To our knowledge, there has only been one other reported case of tibial post fracture with the JI-BCS implant.…”
Section: Discussionsupporting
confidence: 79%
See 1 more Smart Citation
“…On the post's intraoperative evaluation, there were no signs of significant wear due to post impingement. Similar to other post failure reports in deep flexion, the fracture likely occurred due to increased force on the post during squat exercises [ 23 ]. To our knowledge, there has only been one other reported case of tibial post fracture with the JI-BCS implant.…”
Section: Discussionsupporting
confidence: 79%
“…Few cases of tibial post breakage in PS TKA have previously been reported. To our knowledge, 5 previous cases have been reported in the NexGen prosthesis (Zimmer, Warsaw, IN) [ 4 , [19] , [20] , [21] , [22] , [23] , [24] ]. Chiu et al.…”
Section: Discussionmentioning
confidence: 99%