Abstract:).Posterior stabilizing knee prostheses have been increasing in popularity recently due to some studies showing excellent long-term survivorship, more consistent kinematics, technical ease, and higher range of motion.1,2 The posterior stabilized knee relies on a cam and post mechanism to replicate native femoral rollback. One of the major potential complications of this system is post fracture. This occurs in less than 1% of most total knees. 3 These patients usually present with a history of a "pop," effusion… Show more
“…On the post's intraoperative evaluation, there were no signs of significant wear due to post impingement. Similar to other post failure reports in deep flexion, the fracture likely occurred due to increased force on the post during squat exercises [ 23 ]. To our knowledge, there has only been one other reported case of tibial post fracture with the JI-BCS implant.…”
Section: Discussionsupporting
confidence: 79%
“…Few cases of tibial post breakage in PS TKA have previously been reported. To our knowledge, 5 previous cases have been reported in the NexGen prosthesis (Zimmer, Warsaw, IN) [ 4 , [19] , [20] , [21] , [22] , [23] , [24] ]. Chiu et al.…”
Failure of the tibial post in a bicruciate substituting total knee prosthesis is a rare but catastrophic complication. The authors report 2 cases of a fracture of the polyethylene tibial post with subsequent episodes of knee subluxation. Prompt recognition and early revision of these complications are associated with a favorable early outcome.
“…On the post's intraoperative evaluation, there were no signs of significant wear due to post impingement. Similar to other post failure reports in deep flexion, the fracture likely occurred due to increased force on the post during squat exercises [ 23 ]. To our knowledge, there has only been one other reported case of tibial post fracture with the JI-BCS implant.…”
Section: Discussionsupporting
confidence: 79%
“…Few cases of tibial post breakage in PS TKA have previously been reported. To our knowledge, 5 previous cases have been reported in the NexGen prosthesis (Zimmer, Warsaw, IN) [ 4 , [19] , [20] , [21] , [22] , [23] , [24] ]. Chiu et al.…”
Failure of the tibial post in a bicruciate substituting total knee prosthesis is a rare but catastrophic complication. The authors report 2 cases of a fracture of the polyethylene tibial post with subsequent episodes of knee subluxation. Prompt recognition and early revision of these complications are associated with a favorable early outcome.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.