2020
DOI: 10.1101/2020.04.03.20052217
|View full text |Cite
Preprint
|
Sign up to set email alerts
|

Pressure-Regulated Ventilator Splitting (PReVentS) – A COVID-19 Response Paradigm from Yale University

Abstract: In the current COVID-19 crisis, the US and many countries in the world are suffering acute shortages of modern ventilators to care for desperately ill patients. Since modern ICU ventilators are powerful devices that can deliver very high gas flow rates and pressures, multiple physicians have attempted to ventilate more than one patient on a single ventilatorso-called "vent splitting". Early applications of this approach have utilized simple concatenations of ventilator tubing and T-pieces, to provide flow to m… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
17
0

Year Published

2020
2020
2022
2022

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 10 publications
(17 citation statements)
references
References 2 publications
0
17
0
Order By: Relevance
“…Although so-called “co-ventilation” strategies have been used in New York ( 1 ) and have received crisis approval by Health and Human Services and the Federal Drug Administration ( 2 , 3 ), there are major potential risks with such approaches, and major medical organizations have made statements against their use ( 4 ). To address these limitations, several international working groups have designed modified ventilator circuits which permit individualized settings for patients sharing a ventilator ( 5 7 ). These modified systems, also known as differential multiventilation (DMV), use flow restrictors and sensors to allow discrete adjustment and monitoring of inspiratory and expiratory variables (an example system is shown in Fig.…”
mentioning
confidence: 99%
“…Although so-called “co-ventilation” strategies have been used in New York ( 1 ) and have received crisis approval by Health and Human Services and the Federal Drug Administration ( 2 , 3 ), there are major potential risks with such approaches, and major medical organizations have made statements against their use ( 4 ). To address these limitations, several international working groups have designed modified ventilator circuits which permit individualized settings for patients sharing a ventilator ( 5 7 ). These modified systems, also known as differential multiventilation (DMV), use flow restrictors and sensors to allow discrete adjustment and monitoring of inspiratory and expiratory variables (an example system is shown in Fig.…”
mentioning
confidence: 99%
“…Implementation of split breathing circuits without compensating resistances was the model addressed in the joint statement by professional societies recommending against this practice [12]. Various means have been employed empirically to limit flow to the most compliant lungs [17][18][19][20]. Modeling flow restriction adds model parameters, greatly increasing the number of simulations needed.…”
Section: Discussionmentioning
confidence: 99%
“…Some of the assertions of the joint statement are addressed in an appendix [17]. Similar apparatus has been described by an on-line collaborative [18] and another prepublication document from Yale [19]. Clarke et al have reported using simple compression clamps on segments of tubing [20].…”
Section: Introductionmentioning
confidence: 95%
See 1 more Smart Citation
“…A recent non-peer-reviewed report using PCV with pressure-relief valves also implemented a bias flow with a one-way valve to prevent occlusion alarms on the ventilator. 21 Continuous monitoring of arterial oxygen saturation is also desirable in individual patients sharing a single ventilator, for appropriate titration of PEEP, F IO 2 , and detection of acute hypoxic events. Continuous flow monitoring in all connected patients is only necessary for individualization of delivered volume during shared VCV, during which the distribution of volume from the ventilator may suddenly change due to fluctuations in the mechanics of any single patient.…”
Section: Limitationsmentioning
confidence: 99%