Primary UV radiation detector standards

Abstract: Primary standard detectors based on an ionizing chamber and a pneumatic Golay detector have been developed at VNIIOFI to cover the spectral range 30 nm to 400 nm. The double ionizing chamber was used over the spectral range 30 nm to 126 nm and the Golay detector to cover the range from 105 nm to 400 nm. The main advantage of the Golay detector is its high level, non-selective responsivity of 103 V W−1. The responsivity threshold is 10−10 W Hz−1/2. Pulsed synchrotron radiation sources were used for UV beam gene… Show more