2017
DOI: 10.1002/mp.12315
|View full text |Cite
|
Sign up to set email alerts
|

Quantitative ionization chamber alignment to a water surface: Performance of multiple chambers

Abstract: The water-to-air scanning method coupled with gCAP analysis identifies the alignment of the IC OR to the water surface within the scanning resolution for all ICs studied. The gCAP method can precisely and reproducibly align the physical center of a given cylindrical IC with the water surface, be applied prospectively or retrospectively, and provides the prospect for automated water surface identification for scanning systems. The gCAP method eliminates the visual subjectivity inherent to current IC-to-water su… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
6
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(6 citation statements)
references
References 13 publications
0
6
0
Order By: Relevance
“…We used BEAMnrc from the EGSnrc code system to model treatment heads for Varian Cl2100 and Elekta Precise accelerators. These linacs correspond with those used for measurements in our companion paper . The Varian head model from our previously published description has been updated to include electron contributions from various treatment head shielding components, as indicated by other authors .…”
Section: Monte Carlo Modelmentioning
confidence: 99%
See 1 more Smart Citation
“…We used BEAMnrc from the EGSnrc code system to model treatment heads for Varian Cl2100 and Elekta Precise accelerators. These linacs correspond with those used for measurements in our companion paper . The Varian head model from our previously published description has been updated to include electron contributions from various treatment head shielding components, as indicated by other authors .…”
Section: Monte Carlo Modelmentioning
confidence: 99%
“…These linacs correspond with those used for measurements in our companion paper. 17 The Varian head model from our previously published description [18][19][20] has been updated to include electron contributions from various treatment head shielding components, as indicated by other authors. [21][22][23][24] The Elekta Precise head model has been validated and documented in previous publications.…”
Section: Monte Carlo Modelmentioning
confidence: 99%
“…Moreover, it has been stated that the experimental determination of EPOMs is difficult and that a positioning precision of 0.1 mm is necessary to compare different thimble chambers . Positioning detectors at the water surface based on reflection images may introduce systematic errors and the surface tension of the water may obscure the initial depth setting, especially for the plane‐parallel chambers. Therefore, for the plane‐parallel chambers, the initial depth was carefully set and verified with a 3D‐printed, light plastic cone (Fig.…”
Section: Discussionmentioning
confidence: 99%
“…Siebers et al . and Ververs et al proposed calculating the first derivative of the depth dose curve and finding the maximum. This point indicates the position when the wall of a cylindrical detector just reaches the water surface.…”
Section: Discussionmentioning
confidence: 99%
“…To position the IC at the initial depth the most accurate method is to use radiation measurements to localize the surface of the water as the IC is scanned in small steps through the water surface. 10,11 With this method, the surface can be localized with an uncertainty of 0.15 mm.…”
Section: Positioning the Ionization Chamber Along The Beam Axismentioning
confidence: 99%