In fluoroscopy guided interventional procedures, ceiling suspended screen is
used to protect personel for scattered X-ray radiation arising from patents.
The correct positioning of the screen is crucial for proper occupational
radiation protection. The proposed solution in this paper provides reliable
and efficient protection from scattered radiation, based on X-ray device and
the protective screen merging into one system via an appropriate interface.
After the initial manual positioning, automatic repositioning of the screen
is ?xecuted, by curving the screen laterally, clockwise or counterclockwise,
and then with the upper or lower edge forward. All potential clinical
situations were analyzed, considering need for screen position correction:
the semi-automatic solution is designed and realized to follow the medical
procedure in order to keep the efficient level of staff radiation
protection. Furthermore, the assessment of the occupational radiation dose,
provided for screen position optimization, will be imported in the radiation
dose structural report. With application of the universal interface, the
presented solution can be applied not only on newly manufactured ones, but
on existing C-arm X-ray devices as well.