Proceedings of the 2012 International Conference on Quantitative InfraRed Thermography 2012
DOI: 10.21611/qirt.2012.334
|View full text |Cite
|
Sign up to set email alerts
|

Review of current thermal imaging temperature calibration and evaluation facilities, practices and procedures, across EURAMET (European Association of National Metrology Institutes)

Abstract: This paper provides an overview of, and the key findings from, a survey of the thermal imager 'temperature measurement' calibration and evaluation facilities, practices and procedures implemented within the EURAMET region (European Association of National Metrology Institutes, www.euramet.org).

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2014
2014
2022
2022

Publication Types

Select...
2
1
1

Relationship

1
3

Authors

Journals

citations
Cited by 4 publications
(4 citation statements)
references
References 4 publications
0
4
0
Order By: Relevance
“…Other sources of inaccuracies in thermal measurements using TEC-less cameras are [ 46 ]: size-of-source effect, distance effect, and environmental effects, among others, which are also present in the cooled cameras.…”
Section: Introductionmentioning
confidence: 99%
“…Other sources of inaccuracies in thermal measurements using TEC-less cameras are [ 46 ]: size-of-source effect, distance effect, and environmental effects, among others, which are also present in the cooled cameras.…”
Section: Introductionmentioning
confidence: 99%
“…Each thermal imager was calibrated following NPL standard calibration procedure, i.e. in accordance with the procedure usually employed for calibration of single spot radiation thermometers as performed by all EUROMET national measurement institutes (see [2]). The imagers were calibrated using four NPL standard blackbody sources [3] with emissivities >0.999, the characteristics of which can be found in table 2.…”
Section: Calibrationmentioning
confidence: 99%
“…Imager non-uniformity sub-sections 1 This test could not be carried out on imager C as it was unavailable at the time of testing. 2 This test could not be carried out Imager C or D as the measurement region cannot be moved from the central region of the FOV.…”
Section: Figmentioning
confidence: 99%
“…A second problem arises from the dependence of a given pixel’s signal level on stray light or signal levels corresponding to surrounding pixels. This confound is known in the thermal sensing and imaging field as size-of-source effect (SSE){11,12,13,14} but is not assessed in the IRT febrile screening standard{7}. A recent conference publication{10} showed this can cause inner canthi surface temperatures to vary by an order of magnitude more than the surface calibration indicated.…”
Section: Introductionmentioning
confidence: 99%