2019
DOI: 10.32604/mcb.2019.07381
|View full text |Cite
|
Sign up to set email alerts
|

Ring Artifacts Reduction in CBCT: Pixels Detection and Patch Based Correction

Abstract: The ring artifacts introduced by the defective pixels with non-linear responses in the high-resolution detector, have a great impact on subsequent processing and quantitative analysis of the reconstructed images. In this paper, a multistep method is proposed to suppress the ring artifacts of micro CT images, which firstly locates the positions of the defective pixels in the sinogram, and then corrects the corresponding value in the projections. Since the defective pixels always appear as vertical stripes in th… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Publication Types

Select...

Relationship

0
0

Authors

Journals

citations
Cited by 0 publications
references
References 18 publications
0
0
0
Order By: Relevance

No citations

Set email alert for when this publication receives citations?