Runaway pacemaker: A forgotten phenomenon?

Abstract: Runaway is an uncommon pacemaker dysfunction, characterized by fast and erratic spikes at non-physiological rates. This infrequent but potentially lethal failure mode may be related to low battery voltage. Four single chamber pacemaker patients were analyzed (Medtronic Minix ST 8330, Minneapolis, MN, had been implanted in two patients and two CPI Triumph VR 1124, St Paul, MN, in the other two). They had been admitted because of presyncopal episodes. Typical high rate stimuli at 2000 ppm alternating with pacing… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
4
1

Year Published

2008
2008
2023
2023

Publication Types

Select...
5
4
1

Relationship

0
10

Authors

Journals

citations
Cited by 15 publications
(6 citation statements)
references
References 8 publications
0
4
1
Order By: Relevance
“…There are rare reports of erratic behavior of pacemakers at ERI, 16 but we did not observe any erratic or runaway phenomena in our cohort. However, knowledge of this rare phenomenon is important to inform shared decision making.…”
Section: Discussioncontrasting
confidence: 69%
“…There are rare reports of erratic behavior of pacemakers at ERI, 16 but we did not observe any erratic or runaway phenomena in our cohort. However, knowledge of this rare phenomenon is important to inform shared decision making.…”
Section: Discussioncontrasting
confidence: 69%
“…This case demonstrates a unique cause of PoR unrelated to electromagnetic interference and triggered by a series of events leading to violation of the atrial rate limit. Historically, the maximum rate limit was implemented as a safety measure to prevent “runaway pacemaker” syndrome, which is potentially lethal 7 . Through a unique set of physiologic circumstances and device parameters, the maximum atrial rate limit was violated, resulting in a PoR, unipolar maximum output, and pacemaker‐induced extracardiac stimulation of the pectoral muscle (Figure 2D).…”
Section: Discussionmentioning
confidence: 99%
“…The runaway phenomenon is mostly seen in case of a delay in the elective replacement time and excessive decline of battery voltage, particularly in old-generation pacemakers (4). ECG in cases of the typical runaway phenomenon has a slower intermittent ventricular capture or a very-high-rate pacemaker capture, which looks like an artifact.…”
Section: Discussionmentioning
confidence: 99%