Abstract:A solution-based inductively coupled plasma optical emission spectrometric (ICP-OES) method is described for elemental analysis with relative expanded uncertainties on the order of 0.1% relative. The single-element determinations of 64 different elements are presented, with aggregate performance results for the method and parameters for the determination of each element. The performance observed is superior to that previously reported for ICP-OES, resulting from a suite of technical strategies that exploit the… Show more
“…The solutions were gravimetrically diluted to the required concentration. The use of selenium as the internal standard coupled with a bracketing calibration procedure was in line with earlier work that used ICP-OES for high accuracy analysis [11][12][13]. Table 2.…”
Section: Preparation Of Icp-oes Samples and Standardsmentioning
“…The solutions were gravimetrically diluted to the required concentration. The use of selenium as the internal standard coupled with a bracketing calibration procedure was in line with earlier work that used ICP-OES for high accuracy analysis [11][12][13]. Table 2.…”
Section: Preparation Of Icp-oes Samples and Standardsmentioning
“…[13] These direct measurements are done against internationally accepted certified reference materials and shall prove the absence of a significant bias during characterisation of the starting material and production process. Fig.…”
Over the last three years Sigma-Aldrich has built-up a competence centre for CRM (Certified Reference Material) production in Buchs, Switzerland. The new facility is accredited following both ISO/IEC 17025 and ISO Guide 34. The commercial availability of a broad range of different CRMs is of highest importance with respect to the realisation of metrological concepts in practice. This article describes the process of the production and certification of inorganic CRMs of highest quality. This comprises the comprehensive characterisation of the starting materials, the high precision weighing of the compounds, homogenisation and special packaging of these new CRMs.
“…With our lengthy experience in ICP-OES analysis in the metrological measurement environment, we accurately measure inorganic P as demonstrated by one of our peer metrology institutes [22]. To fully realize this excellent accuracy of ICP-OES in determination of P from the target nucleotide, extreme care has been taken.…”
Section: Quantitation Of P Using Icp-oesmentioning
confidence: 99%
“…The number of moles of P atoms determined by this procedure is converted to the number of moles of the oligonucleotide by dividing it with the number of phosphodiester bonds in the oligonucleotide. Highly elemental speciWc and reproducible ICP-OES analysis combined with rigorous drift correction permits the determination of major inorganic constituents in solution with an uncertainty of only 0.1% [22,23]. The concept of the proposed analytical approach (Fig.…”
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.