Purpose: The association between the extent of the wire and device bias as assessed by optical coherence tomography (OCT) in the healthy portion of the vessel and the risk of coronary artery injury after orbital atherectomy (OA) has not been fully elucidated. Thus, purpose of this study is to investigate the association between pre-OA OCT findings and post-OA coronary artery injury by OCT. Methods: We enrolled 148 de novo lesions having calcified lesion required OA (max Ca angle > 90°) in 135 patients who underwent both pre- and post-OA OCT. In pre-OA OCT, OCT catheter contact angle and the presence or absences of guide-wire (GW) contact with the normal vessel intima were assessed. Also, in post-OA OCT, we assessed there was post-OA coronary artery injury (OA injury), defined as disappearance of both of intima and medial wall of normal vessel, or not. Results: OA injury was found in 19 lesions (13%). Pre-PCI OCT catheter contact angle with the normal coronary artery was significantly larger (median 137°; inter quartile range [IQR] 113–169 vs. median 0°; IQR 0–0, P < 0.001) and more GW contact with the normal vessel was found (63% vs. 8%, P < 0.001). Pre-PCI OCT catheter contact angle > 92° and GW contact with the normal vessel intima were associated with post-OA vascular injury (Both: 92% (11/12), Either: 32% (8/25), Neither: 0% (0/111), P < 0.001). Conclusion: Pre-PCI OCT findings, such as catheter contact angle > 92° and guide-wire contact to the normal coronary artery, were associated with post-OA coronary artery injury.