Sub-millimeter endoscope demonstrates feasibility of in vivo reflectance imaging, fluorescence imaging, and cell collection in the fallopian tubes

Abstract: . Significance: Most cases of high-grade serous ovarian carcinoma originate as serous tubal intraepithelial carcinoma (STIC) lesions in the fallopian tube epithelium (FTE), enabling early endoscopic detection. Aim: The cell-acquiring fallopian endoscope (CAFE) was built to meet requirements for locating potentially pathological tissue indicated by an alteration in autofluorescence or presence of a targeted fluorophore. A channel was included for directed scrape b… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0

Year Published

2022
2022
2025
2025

Publication Types

Select...
4
2
1

Relationship

2
5

Authors

Journals

citations
Cited by 14 publications
(6 citation statements)
references
References 45 publications
0
5
0
Order By: Relevance
“…However, a more robust solution may be to eliminate the ferrule by bonding the fibers, distal tip optics, and pull wires directly to the PEEK MLE material. Glass-PEEK bonding can be accomplished with Loctite ® 4013 cyanoacrylate liquid adhesive; 23 however, SS pull wires would likely require a non-adhesive method of distal fixation.…”
Section: Pilot Study Lessons Learnedmentioning
confidence: 99%
“…However, a more robust solution may be to eliminate the ferrule by bonding the fibers, distal tip optics, and pull wires directly to the PEEK MLE material. Glass-PEEK bonding can be accomplished with Loctite ® 4013 cyanoacrylate liquid adhesive; 23 however, SS pull wires would likely require a non-adhesive method of distal fixation.…”
Section: Pilot Study Lessons Learnedmentioning
confidence: 99%
“…The previous CAFE endoscope used a nitinol wire with a distal hook shape, which was extended from the endoscope, spun to scrape the fallopian tube epithelium, and could collect on the order of 10 5 cells per scrape biopsy. 29 However, there were issues with its retractability during operation, and potential for harm to the tissue caused by the sharp wire tip. This is the motivation behind the new approach, the scoop method.…”
Section: Cell Collection Improvementsmentioning
confidence: 99%
“…The resolution was greatly improved, now able to resolve down to Group 5 Element 6 (8.77 µm), instead of Group 2 Element 4 (88 µm) as seen with the previous system. 29…”
Section: Imaging System Performancementioning
confidence: 99%
See 2 more Smart Citations