Abstract:This case describes some of the commonly overlooked device-related issues in patients who have reportedly failed to respond to cardiac resynchronization therapy (CRT). The case demonstrates voltage-dependent right ventricular capture instead of right atrial capture by a subtly malpositioned right atrial lead. CRT therapy failed to improve symptoms of heart failure and the diagnosis of "CRT nonresponder" was made. With a detailed fact-finding approach, the mechanism behind this nonresponse was identified, and t… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.