2015
DOI: 10.1111/jce.12867
|View full text |Cite
|
Sign up to set email alerts
|

Sudden Cardiac Death Despite a Functional Cardioverter‐Defibrillator: The Case for Early and Aggressive Therapy for Ventricular Tachycardia in Selected Patients

Abstract: We present three cases within 11 months at a single institution of sustained VT that fell below the programmed detection rate of the patients' implantable cardioverter-defibrillators (ICDs), two of which continued until converting to an agonal VF that did not meet criteria for detection, and a third case that could not be successfully defibrillated after a prolonged period of VT. These episodes may be under-recognized due to the dependence of device diagnostic storage on programming and the post-mortem effort … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
6
0

Year Published

2016
2016
2022
2022

Publication Types

Select...
5
1
1

Relationship

0
7

Authors

Journals

citations
Cited by 7 publications
(6 citation statements)
references
References 14 publications
0
6
0
Order By: Relevance
“…Although low-frequency VF has been considered an agonal rhythm, metabolic changes during VT may produce de novo, low-frequency VF (patients 3, 7, and 9) because of local conduction block or hyperkalemia, resulting in long device-detected intervals because of undersensing or entrance block. 6 Patient 5’s first recorded rhythm was low-frequency VF, but recurrence of high-frequency VF promptly after the first defibrillation indicates that the presenting VF was not an agonal rhythm. In addition, prolonged VT may cause metabolic changes that facilitate early postshock recurrence of VT/VF (eg, catecholamine release, patients 4, 5, 7–9).…”
Section: Discussionmentioning
confidence: 98%
See 1 more Smart Citation
“…Although low-frequency VF has been considered an agonal rhythm, metabolic changes during VT may produce de novo, low-frequency VF (patients 3, 7, and 9) because of local conduction block or hyperkalemia, resulting in long device-detected intervals because of undersensing or entrance block. 6 Patient 5’s first recorded rhythm was low-frequency VF, but recurrence of high-frequency VF promptly after the first defibrillation indicates that the presenting VF was not an agonal rhythm. In addition, prolonged VT may cause metabolic changes that facilitate early postshock recurrence of VT/VF (eg, catecholamine release, patients 4, 5, 7–9).…”
Section: Discussionmentioning
confidence: 98%
“…Manufacturers responded with improved technology; in this century, reports of failure to treat life-threatening VT or VF have been rare and limited to one 1 4 or a few patients. 5 , 6 …”
mentioning
confidence: 99%
“…Montgomery et al. reported on three cases of sudden death due to sustained VT that fell below the programmed detection rate . Anguera et al.…”
Section: Discussionmentioning
confidence: 99%
“…A matter of concern is the anecdotal reports of sudden cardiac death despite a functional defibrillator . In this report, we describe another 24 patients with a fatal or near‐fatal arrhythmic event due to failure to detect ongoing life‐threatening ventricular tachyarrhythmias (VAs).…”
Section: Introductionmentioning
confidence: 98%
See 1 more Smart Citation