Background: While conventional multiparameter flow cytometers have proven highly successful, there are several types of analytical measurements that would benefit from a more comprehensive and flexible approach to spectral analysis including, but certainly not limited to spectral deconvolution of overlapping emission spectra, fluorescence resonance energy transfer measurements, metachromic dye analysis, free versus bound dye resolution, and Raman spectroscopy. Methods: Our system utilizes a diffraction grating to disperse the collected fluorescence and side-scattered light from cells or microspheres passing through the interrogation region over a rectangular charge-coupled-device image sensor. The flow cell and collection optics are taken from a conventional flow cytometer with minimal modifications to assure modularity of the system. Results: Calibration of the prototype spectral analysis flow cytometer included wavelength characterization and calibration of the dispersive optics. Benchmarking of the system demonstrated a single particle/cell intensity sensitivity of 2160 MESF of R-Phycoerythrin. Single particle spectra taken with our instrument were validated against bulk solution fluorimeter and conventional flow cytometer measurements. Coefficients of variation of integrated