The development of selected reaction monitoring methods for targeted proteomics via empirical refinement

Abstract: Software advancements in the last several years have had a significant impact on proteomics from method development to data analysis. Herein we detail a method, which uses our in-house developed software tool termed Skyline, for empirical refinement of candidate peptides from targeted proteins. The method consists of 4 main steps from generation of a testable hypothesis, method development, peptide refinement, to peptide validation. The ultimate goal is to identify the best performing peptide in terms of ioniz… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
88
0

Year Published

2013
2013
2022
2022

Publication Types

Select...
7
1

Relationship

2
6

Authors

Journals

citations
Cited by 105 publications
(88 citation statements)
references
References 55 publications
0
88
0
Order By: Relevance
“…The result files of the MaxQuant search obtained from the shotgun analyses were used to build an MS/MS library in Skyline, version 2.5 [21,22]. Besides the proteins selected for quality control assessment, as described in the BResultsŝ ection, 31 regulated proteins were considered for MRM analysis.…”
Section: Mrm Assay Developmentmentioning
confidence: 99%
“…The result files of the MaxQuant search obtained from the shotgun analyses were used to build an MS/MS library in Skyline, version 2.5 [21,22]. Besides the proteins selected for quality control assessment, as described in the BResultsŝ ection, 31 regulated proteins were considered for MRM analysis.…”
Section: Mrm Assay Developmentmentioning
confidence: 99%
“…For the QTRAP 5500 all transitions were selected to be below m/z 1,250 due to the instrument's upper mass limit of m/z 1250 in Q1 and Q3. For the ThermoFisher Scientific and Agilent platforms, optimal peptides and their corresponding transitions were determined by predicting all tryptic peptides from the 6 ProteinMix-QC and by monitoring all y-ions from y 3 to y n-1 on the triple quadrupole TSQ Quantum-Ultra and Agilent 6460 (29). The raw data were imported into Skyline and results were refined to select the 5 most abundant transitions for all detectable peptides.…”
Section: Methodsmentioning
confidence: 99%
“…Skyline was used to build spectral libraries from data dependent acquisitions and peptide search engine results using the BiblioSpec library builder (28) so that MRM transitions could be selected based on previously acquired discovery platform data (for AB Sciex and Waters instruments). Alternatively (for Thermo and Agilent instruments), SRM Refinement approaches (29) were used on a triple quadrupole MS to determine optimal MRM transitions for each peptide.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…A variety of methods have been previously used to identify signature peptides for protein quantification (2,15,16 ). One common approach is to target peptides identified by a data-dependent MS screen on related samples, as these peptides are guaranteed to be detectable by MS (2,17 ).…”
Section: © 2015 American Association For Clinical Chemistrymentioning