The effect of adjusting the stroke value of jet dispenser on the performance of glucose biosensor with gold-plated electrode

Abstract: This study investigates the effect of dispensing glucose oxidase enzyme onto the reaction zone of a golden electrode via a jetting dispenser. Each droplet of enzyme solution dispensed onto the reaction zone of golden electrode weighs exactly the same at around 0.4 mg. This study shows that the spring and needle assembly can be controlled by adjusting the stroke value of the stroke adjustment knob to generate different jet dispensing effects, thus affecting the change of droplets of glucose oxidase enzyme solut… Show more