1984
DOI: 10.1111/j.1540-8159.1984.tb05685.x
|View full text |Cite
|
Sign up to set email alerts
|

The Importance of Software Programmable Pacemakers: In Vivo Programming of a Prototype Device

Abstract: Recently, we have developed microcomputer-oriented pacemakers with true software flexibility. The first 25 devices implanted (DPG-1 type) exhibited an unexpected phenomenon caused by a microcomputer error. A safety mechanism switched off all special functions and automatically reset all parameters to standard settings. The problem was solved by noninvasive modification of the software. More than 200 units (TX-1 type), which are rate responsive pacemakers based on the QT principal, were implanted up to May 1984… Show more