The international calibration procedure for B3 film dosimetry system to ensure the quality irradiated products by 10 MeV electron beam accelerators at VINAGAMMA
Abstract:We performed an in-plant calibration of dosimetry system by electron beam (EB)irradiating the B3 film dosimeters at VINAGAMMA, and inter-compared with the alanine dosimetry, which were supplied and analyzed by Risø High Dose Reference Laboratory (HDRL) as the reference standard. The results revealed that the relative deviation between the values of absorbed doses obtained with our dosimeter and the transfer standards dosimeter measured by HDRL was within the acceptable limitation (about ± 3.0 % in the target r… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.