The international calibration procedure for B3 film dosimetry system to ensure the quality irradiated products by 10 MeV electron beam accelerators at VINAGAMMA

Abstract: We performed an in-plant calibration of dosimetry system by electron beam (EB)irradiating the B3 film dosimeters at VINAGAMMA, and inter-compared with the alanine dosimetry, which were supplied and analyzed by Risø High Dose Reference Laboratory (HDRL) as the reference standard. The results revealed that the relative deviation between the values of absorbed doses obtained with our dosimeter and the transfer standards dosimeter measured by HDRL was within the acceptable limitation (about ± 3.0 % in the target r… Show more