Thermography-Based Comparison of Irrigation Temperatures Between Moses Mode and Virtual Basket Mode: An In Vitro Phantom Study

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
3

Relationship

2
1

Authors

Journals

citations
Cited by 3 publications
(6 citation statements)
references
References 20 publications
0
6
0
Order By: Relevance
“…The ratios of the high-temperature areas in the LPM and MM tended to be higher than those in the SPM and VBM. These results are understandable considering that the time to reach a threshold in the LPM and MM is significantly shorter than that in the SPM and VBM [ 5 ]. However, the exact etiology of the differences in heat production between different laser pulse modes remains unclear.…”
Section: Discussionmentioning
confidence: 97%
See 2 more Smart Citations
“…The ratios of the high-temperature areas in the LPM and MM tended to be higher than those in the SPM and VBM. These results are understandable considering that the time to reach a threshold in the LPM and MM is significantly shorter than that in the SPM and VBM [ 5 ]. However, the exact etiology of the differences in heat production between different laser pulse modes remains unclear.…”
Section: Discussionmentioning
confidence: 97%
“…Winship et al reported that the MM produces significantly lower temperatures compared with the LPM [ 4 ]. In addition, the irrigation temperature in the MM could rise more rapidly than that in the VBM [ 5 ]. However, the influence of these laser pulse modes on dynamic changes in the area or the spread of high temperatures remains unclear.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…An artificial kidney model was created by using Urovac evacuator™ (Boston Scientific, Marlborough, MA, USA; capacity: 340 ml, length: 14 cm, maximum diameter: 7.3 cm, thickness: 3.2 mm) as a renal pelvis calyx based on previous studies [ 12 , 13 ]. The nephrostomy sheath was connected to the Urovac evacuator™ so that it was 70° above the horizontal plane, which was designed to represent puncture of the lower calyx in a prone position in a human body.…”
Section: Methodsmentioning
confidence: 99%
“…For this reason, several factors might be used by the operators to keep the temperature in a safe profile [39]. With irrigation, cooler water circulates through the system, dissipating heat [40,41]. The use of a larger calibre ureteral access sheath (UAS) (12/14 F or 14/16 Fr) results in temperatures around 43 • C, and in improved irrigation and lower intrarenal pressures [39,40].…”
Section: Role Of Temperature With Laser Usementioning